Standardizing wrist x-rays: The human-centered design and evaluation of a wrist x-ray positioning tool

Laura Vancer, Maxwell Campbell, Glenn Schurmans, John Henry, Jamie Kueneman, Joy MacDermid, Gregory Garvin, Emily Lalone

Abstract

Introduction/Background

Small amounts of malrotation can significantly decrease the diagnostic utility of a wrist radiograph. Poorly positioned wrist radiographs can result in repeat scans, additional imaging, and missed injuries. Tolerance for mistakes must be low. Radiographers, especially those with limited training, experience, or resources, may struggle to meet an acceptable standard, underscoring the need for an assistive positioning device that makes proper positioning more accessible.

Introduction

Wrist x-ray positioning

Wrists are one of the most commonly injured anatomical regions, accounting for approximately 20% of all hospitalized fractures in Canada [1]. A wrist fracture or forced hyperextension sprain can be associated with a ruptured scapholunate (SL) ligament [2] in 32% of individuals. SL rupture, also known as SL dissociation, can lead to a form of wrist instability called SL advanced collapse (SLAC)—which is responsible for approximately 55% of wrist osteoarthritis [3]. While osteoarthritis is the most common degenerative joint disease affecting individuals over the age of 18, there are currently no effective therapeutic treatments [4]. 

Methods

The HCD process was implemented during the creation of the Tru-Vu through the three IDEO HCD stages: inspiration, ideation, and implementation [22]. Each stage consisted of varying methods (Fig 1), with consistent communication with end-users throughout the entirety of the design process.

Results

Throughout each stage of the design process—inspiration, ideation, and implementation—the device underwent multiple iterations, guided by end-user feedback.

Inspiration

Preliminary interviews revealed overarching themes relating to the quality of wrist radiographs and factors that influence the adoption of novel positioning technologies. Six primary factors affecting radiograph quality were articulated during interviews: radiographer experience, practice setting (rural/small-town versus urban hospital centers), differences between educational training and clinical practice, the complexity of carpal anatomy, understanding of diagnostic objectives, and challenges related to patient positioning. 

Discussion

This study followed the human-centered design (HCD) process to develop the novel Tru-Vu positioning aid and training modules. Wrist radiographs have relatively high rejection rates and due to the limitations of current solutions, many radiographers rely on their best judgement to position the wrist without the use of assistive devices. Through an iterative design process with continuous feedback from end-users, the Tru-Vu tool and training modules were developed to standardize wrist positioning and as a result improve the diagnostic quality of common wrist radiographs.

Conclusion

The development of the novel Tru-Vu positioning aid and training modules used a HCD approach in response to the challenge of capturing high-quality wrist radiographs. Throughout the three design stages—inspiration, ideation, and implementation—continuous collaboration with a team of radiologists, orthopedic surgeons, and radiographers provided the team with a holistic understanding of the initial challenge and the impact of different design decisions along the way.

Citation: Vancer L, Campbell M, Schurmans G, Henry J, Kueneman J, MacDermid J, et al. (2026) Standardizing wrist x-rays: The human-centered design and evaluation of a wrist x-ray positioning tool. PLoS One 21(10): e0359437. https://doi.org/10.1371/journal.pone.0359437

Editor: Yi Kai Wong, IMU University, MALAYSIA

Received: October 16, 2025; Accepted: September 14, 2026; Published: October 5, 2026

Copyright: © 2026 Vancer et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

Data Availability: All relevant data are within the manuscript and its Supporting information files.

Funding: This study was funded through an Innovation Grant from the Arthritis Society of Canada (IG-21-0000000017) and a Frugal Biomedical Innovations Catalyst Grant from Western University, received by EL. LV completed this work with funding from the Ontario Graduate Scholarship and an NSERC CREATE in Engineering Health Equity scholarship. The funders did not play a role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript.

Competing interests: The authors acknowledge the following competing interests: Authors Emily Lalone and Maxwell Campbell are inventors on a Canadian patent (No. 3,271,804) and a US patent (19/189,513) that are currently pending.

 

https://journals.plos.org/plosone/article?id=10.1371/journal.pone.0359437